Bellovac ABT Скачать руководство пользователя страница 1

Open the package using aseptic pro-

cedures.

Place the catheter in situ. Remove 

the trocar, secure the catheter, and 

connect the catheter to the inlet

tubing.

1. To vacuum the system and 

evacuate air from the operation

area, close the inlet clamp.

2. Compress the bellows.

3. Open the inlet clamp again. 

Repeat steps 1–3 until blood

starts to flow, or the bellows

remain in a compressed, 

vacuumed state.

4. Close the outlet clamp.

If tourniquet is used: To avoid

clotting, close the inlet clamp

after the last compression of the

bellows. Reopen the inlet clamp

to start collection approximately

15 minutes after the release of

the tourniquet.

5. Remove the protective cap from

the bag inlet.

6. Attach the Autotransfusion bag to

the outlet port of the bellows

using the screw connector. 

7. Mark with the identity of the

patient, what time the collection

started, and definite time when the

collection should be terminated on

the bag (blood collection should

be completed within 6 hours of

the start of collection).

Hang the Bellovac

ABT in an upright

position, using the bed rail strap.

Send the Transfusion Set, with the

patient, to the postoperative ward.

INSTRUCTIONS FOR USE

SET UP

1

2

3

4

5

6

7