The E-Line 80 is a dual scale Brix refractometer ideal for use in the 
food industry and other applications requiring concentration 

measurements over a wide measurement range. 

 

Applying the sample to the refractometer

Lift the illuminating prism

1

 and apply the sample to the measurement 

prism

2

.  Close the prism and view the borderline (light/dark 

demarcation line) through the eyepiece

3

.  

 

Taking a reading: 

Depending on the concentration of the sample, select the range of 
measurement

(Table 1)

 by adjusting the scale selection control

4

.  If the 

concentration is unknown, apply the sample and step through the 

ranges until a clear borderline is found.  

 

 

The concentration of the sample can be measured at the point the 

borderline crosses the scale.  Take care to ensure that the reading is 
taken from the correct scale range.   

 

Calibration: 

The E-line 80 is calibrated to read accurately at 20

˚

C.  Readings may 

be taken at any other temperature by applying a correction factor 
(Table 2).   

 

Alternatively the E-line 80 has the facility to correct for temperature by 
manual offset.  Prior to taking a reading, ensure the temperature of the 

instrument is stable.  Apply distilled water and check that the reading 
is zero (Scale 1).  If the reading is not zero, use the screwdriver 

supplied to adjust the zero calibration screw 

so that the borderline 

crosses the zero mark on the scale.    No temperature correction is 
required for this method but as the relationship between refractive 

index and temperature is not linear, small errors may occur at the top 
end of the scale compared to readings near water.   

 

 

 

 

     

     

 

 

Always check sample Health & Safety Data before applying to the 
refractometer. 

 

When applying samples to the prism which are likely to cause harm to 
skin or eyes, wear appropriate protective clothing and glasses. 

 

This refractometer is a precision optical instrument and should be 
handled with care.   Do not drop or subject it to sharp knocks. 

 

Optical glass is relatively soft and care should be taken not to scratch 
the prism surface.   Do not use metal spatulas or glass rods to apply 

samples but instead use softer materials such as plastic. 

 

 

 

 

 

Table 1 

Bell Stanley Ltd.

Longfield Road, Tunbridge Wells, Kent TN2 3EY, United Kingdom. 
Telephone: +44 (0) 1892 500400   Fax: +44 (0) 1892 543115 
Website: 

www.bellinghamandstanley.com

  Email: [email protected] 

Specification

 

Range 
    Scale One 

    Scale Two     

0-80 

˚

Brix 

0-50 

˚

Brix 

50-80 

˚

Brix 

Minimum Division 

0.1 

˚

Brix 

E-Line 80 Refractometer

Instruction Manual

Warning

Order Code Description

 

44-807 

E-Line 80 Refractometer 

44-893  Iss 01 Rev 0 

E-Line 80 Refractometer Illuminating Prism closed 

E-Line 80 Refractometer Illuminating Prism open 

Eyepiece

3

 

Illuminating 

Prism

1

 

Scale Selection Control

4

 

Zero Calibration 

Screw

5

 

Measurement 

Prism

2

 