PX106545-EN, Rev.A
43
6.5
CALIBRATION
To display splice mode, press
and select the “CALIBRATION” menu
with the
button. The calibration menu is equipped with various functions, such as arc amount
calibration, motor operation test, etc.
6.5.1
Arc calibration
The FX Fusion Splicer continuously checks for a change in temperature and air pressure
through each sensor. Based on such data, arc amount is automatically adjusted. A change
in arc amount due to abrasion of the electrodes or the fiber splice, however, is not
automatically adjusted. The central axis of the arc can also be moved toward the left or right.
In this case, arc calibration is required.
When executing arc calibration, arc voltage is automatically changed to a
proper value. This value is calculated internally; the arc voltage cannot
arbitrarily be changed.
Only SM fiber should be used for arc calibration.