Operation BDL
58
METPOINT® BDL
14.2.3.3 Modbus settings for the METPOINT
®
SD23
When connecting the METPOINT
®
SD23 via Modbus, the following settings are required:
First step:
select a free sensor channel
Main menu
Settings
Sensor settings
Select a free channel (example: channel A1)
Second step:
select Modbus type
Main menu
Settings
Sensor settings
A1
Type text field
Select Modbus and confirm with
>OK<.
Third step:
define a name
Main menu
Settings
Sensor settings
A1
Name text field
Now, a
name
needs to be entered.
Fourth step:
define the Modbus settings
Main menu
Settings
Sensor settings
A1
Modbus settings
Note:
Further information regarding "Marking and setting text fields" can be found in Chapter
Verweisquelle konnte nicht gefunden werden.
The corresponding Modbus ID can be taken
from the data sheet of the sensor (here, for
example 1).
Adjust the other parameters according to the
illustration.