Error message in setup mode
Error
Description
(ENTRY)
(DENIED)
You have tried to change/set a parameter in SAFE set-up, which is
only possible in FULL set-up.
(LIN.PT
(LO)
Trial to linearization underneath the zero point
(PT.TOO)
(CLOSE)
Adjustment points are too close to one another (<2%) Adjustment
impossible. Use other (usually higher) weights.
(RES)
(LO)
Too little calibration values (<100) set.
(RES)
(HIGH)
Too many calibration values set.
(SPAN)
(LO)
Weighing cell signal too low for this setting
(SPAN)
(HIGH)
Weighing cell signal too high for this setting or will become too
high in case of full load.
(ZERO)
(LO)
Zero point below -2mV/V
(ZERO)
(HIGH)
Zero point above +2mV/V