Beijing Choice Electronic Technology MD300C1C Скачать руководство пользователя страница 1

OPERATOR’S 

MANUAL 

V4.0C1 

Fingertip Pulse Oximeter

 

 

General Description   

Oxygen  Saturation  is  a  percentage  of  Oxyhemoglobin  (HbO2)  capacity,  compounded 
with oxygen, by all combinative hemoglobin (Hb) capacity in blood. In other words, it is 
consistency  of  Oxyhemoglobin  in  blood.  It  is  a  very  important  parameter  for  the 
Respiratory  Circulation  System.  Many  respiratory  diseases  can  result  in  oxygen 
saturation being lowered in human blood. Additionally, the following factors can reduce 
oxygen  saturation:  Automatic  regulation  of  organ  dysfunction  caused  by  Anesthesia, 
Intensive  Postoperative  Trauma,  injuries  caused  by  some  medical  examinations.  That 
situation might result in light-headedness, asthenia, and vomiting. Therefore, it is very 
important to know the oxygen saturation of a patient so that doctors can find problems in 
a timely manner. 

The  fingertip  pulse  Oximeter  features  small  size,  low  power  consumption,  convenient 
operation and portability. It is only necessary for a patient to put one of his fingers into 
the fingertip photoelectric sensor for diagnosis, and a display screen will show oxygen 
saturation. It has been proven in clinical experiments that it also features high precision 
and repeatability. 

 

Measurement principle 

Principle  of  the  Oximeter  is  as  follows:  A  mathematical  formula is  established  making 
use  of  Lambert  Beer  Law  according  to  Spectrum  Absorption  Characteristics  of 
Reductive  hemoglobin(RHb)  and  Oxyhemoglobin  (HbO2)  in  glow  and  near-infrared 
zones.  Operation  principle  of  the  instrument:  Photoelectric  Oxyhemoglobin  Inspection 
Technology  is  adopted  in  accordance  with  Capacity  Pulse  Scanning  and  Recording 
Technology, so that two beams of different wavelength of lights (660nm glow and 940nm 
near infrared light) can be focused onto a human nail tip through a clamping finger-type 
sensor. A measured signal obtained by a photosensitive element, will be shown on  the 
Oximeter

’s display through process in electronic circuits and microprocessor shown on 

the Oximeter

’s display through electronic circuits and a microprocessor. 

Diagram of Operation Principle 

1. Red and Infrared-ray Emission Tube 
2. Red and Infrared-ray Receipt Tube 

Precautions for use 

Do not use the pulse oximeter in an MRI or CT environment 

Do not  use the  pulse oximeter in situations  where alarms are required. The device 
has no alarms. 

3  Explosion hazard:

 Do not use the pulse oximeter in an explosive atmosphere. 

The pulse oximeter is intended only as an adjunct in patient assessment. It must be 
used in conjunction with other methods of assessing clinical signs and symptoms. 

Check  the  pulse  oximeter  sensor  application  site 

frequently

  to  determine  the 

positioning of the sensor and circulation and skin sensitivity of the patient.   

Do not stretch the adhesive tape while applying the pulse oximeter sensor. This may 
cause inaccurate readings or skin blisters. 

Before use, carefully read the manual. 

The pulse oximeter has no SpO

2

 alarms; it is not for continuous monitoring.   

Prolonged  use  or  the  patient’s  condition  may  require  changing  the  sensor  site 
periodically.  Change  sensor  site  and  check  skin  integrity,  circulatory  status,  and 
correct alignment at least every 4 hours. 

10 

Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, 
or immersing the sensors in liquid. 

11 

Significant  levels  of  dysfunctional  hemoglobins  (such  as  carbonxy-  hemoglobin  or 
methemoglobin) may affect the readings. 

12 

Intravascular dyes such as indocyanine green or methylene blue 

13 

SpO

2

 measurements may be adversely affected in the presence of high ambient light. 

Shield  the  sensor  area

 

(with  a  surgical  towel,  or  direct  sunlight,  for  example)  if 

necessary. 

14 

Excessive patient movement may cause inaccurate readings. 

15 

High-frequency electrosurgical interference may cause inaccurate readings. 

16 

Venous pulsations may cause inaccurate readings. 

17 

Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or 
intravascular line. 

18 

The  patient  has  hypotension,  severe  vasoconstriction,  severe  anemia,  or 
hypothermia. 

19 

The patient is in cardiac arrest or is in shock. 

20 

Fingernail polish or false fingernails may cause inaccurate SpO

2

 readings. 

Follow  local  ordinances  and  recycling  instructions  regarding  disposal  or 
recycling of the device and device components, including batteries. 

Product Properties 

Operation of the product is simple and convenient. 

The product is small in volume, light in weight and convenient in carrying. 

Power  consumption  of  the  product  is  low  and  the  two  AAA  batteries  can  be 
operated continuously for 30 hours.   

A low voltage warning will be indicated in visual window when battery voltage is 
so low that normal operation of the oximeter might be influenced. 

The product will automatically be powered off when no signal is in the product for 
longer than 8 seconds. 

Product Operation Scope 

Fingertip PULSE OXIMETER is a portable non-invasive, spot-check, oxygen saturation 
of arterial hemoglobin (SpO

2

) and pulse rate of adult and pediatric patient at home, and 

hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc).  It is 
not for continuously monitoring. 

The  PULSE  OXIMETER  requires  no  routine  calibration  or  maintenance  other  than 
replacement of batteries. 

Operation Instructions 

Install two AAA batteries into battery cassette before closing its cover. 

Place clamp over finger nail as the following diagram. 

Insert one finger into rubber hole of the Oximeter fully. 

Press the switch for one time on front panel. 

Finger and body should not tremble during measuring. 

Read corresponding data on the display screen. 

Note:

 Please use medical alcohol to clean the rubber, inside the Oximeter, that touches 

the finger. Also, clean the test finger with alcohol before and after each test. (The rubber 
inside of the Oximeter is medical rubber, which has no toxins, and is not harmful to the 
skin). 

When  your  finger  is  inserted  into  the  Oximeter,  your  nail 
surface must be upward.

 

Operation Instructions 

 

The Pulse bar graph displays corresponding with the patient

’s pulse beat. The height of 

the bar graph shows the patient

’s pulse strength. 

 

Product Accessories 

1. 

One strap 

2. 

Two batteries 

3. 

One user

’s manual

 

Battery Installation 

1. 

Put the two AAA batteries into battery compartment in correct polarities. 

2. 

Push the battery cover horizontally along the arrow shown as below: 

Notes:   

 

Battery polarities should be correctly installed. 
Otherwise,  damage  may  be  caused  to  the 
device. 

 

Please  put  in  or  remove  batteries  in  right 
order,  or  may  cause  damage  to  the  device 
bracket. 

 

Please  remove  the  batteries  if  the  Oximeter 
will not be used for a long time.

 

 

Strap Installation 

1. 

Thread thinner end of the strap through the loop. 

2. 

Thread thicker end of the strap through the threaded end before pulling it tightly. 

Calibrating the pulse oximeter 

1. 

The functional tester cannot be used to assess the accuracy of the oximeter. 

2. 

The  test  methods  used  to  establish  the  SpO

2

  accuracy  is  clinical  testing.  The 

oximeter used to measure the arterial haemoglobin oxygen saturation levels and 
these  levels  are  to  be  compared  to  the  levels  determined  from  arterial  blood 
sampling with a CO-oximeter. 

3. 

Index 2 made by Bioteck company is a function tester. Set Tech to 1, R curve to 2, 
then user can use this particular calibration curve to measure the oximeter. 

MD300C1 

Отзывы: