16
Sensor change overdue
reminder
This reminder is displayed at
switch on when the sensor
requires replacing. This
reminder cannot be bypassed.
Maintenance
Routine maintenance
1.
Mouthpieces should be replaced after every use.
2.
Hands should be washed regularly in accordance with infection control practice.
3.
Please do not attempt to modify the equipment in any way or use accessories not specified
by the manufacturer. Any attempt to do so will invalidate the warranty and may
compromise the safety of the device.
4.
Bedfont® will make available upon request service training to appropriately qualified
persons.
5.
The Micro
+
™ Smokerlyzer® requires calibrating ever
y 6 months.
6.
Holding the reset button down for 30 seconds will perform complete device reset,
this will
clear any saved data and revert all settings to the factory defaults.
After performing a reset
the device will need to have the date/time set and be calibrated before it can be used.
7.
Replace batteries when indicated by the empty symbol.
8.
Bedfont® recommend removal of the batteries when the device is not used for prolonged
periods of time to prevent leakage.
9.
Replace breath sampling D-piece
™
every 30 days or if visibly soiled or contaminated. The
Micro
+
™ Smokerlyzer®
will give a reminder during start-up when the D-piece
™
should be
replaced,
see ‘
change D-piece
™’ symbol.
10.
The sensor should be replaced every 5 years, 60 days prior to the sensor
change ‘countdown
to sensor change’ symbol will be shown with the date on which the sensor should be
changed. This can be ignored by pressing the home icon until the date at which the sensor
should be changed arrives, at this point the ‘change sensor’ symb
ol will be shown. Change
sensor if trained to do so by an approved Bedfont® engineer or send to Bedfont® or the local
representative.