57
Emissions
The FeNObreath™ Monitor is intended for use in the electromagnetic environment specified below. The user should
assure that it is used in such an environment.
Emission Tests
Compliance
Electromagnetic Environment –
Guidance
RF emissions CISPR 11
Group 1
The FeNObreath™ Monitor uses RF
energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11
Class A
The FeNObreath™ Monitor is suitable for
use in all establishments other than
domestic, and may be used in domestic
establishments and those directly
connected to the public low- voltage
power supply network that supplies
buildings used for domestic purposes,
provided the following warning is
heeded:
WARNING:
This equipment/system is
intended for use by healthcare
professionals only. This equipment/
system may cause radio interference or
may disrupt the operation of nearby
equipment. It may be necessary to take
mitigation measures, such as re-orienting
or relocating the FeNObreath™ monitor
or shielding the location.
Immunity
Guidance and manufacturer’s declaration: Electromagnetic immunity
The FeNObreath™ is intended for use in the electromagnetic environment specified below. The customer or the user of
the FeNObreath™ should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
guidance
Electrostatic Discharge
(ESD) IEC 61000-4-2
±8kV contact
± 2, 4, 8 and 15kV air
±8kV contact
± 2, 4, 8 and 15kV air
Floor should be wood, concrete or
ceramic floor tile. If floors are
covered with synthetic material the
relative humidity should be at least