53
Glossary of Symbols and Safety Information
Glossary of Symbols
Title of Symbol
Symbol
Explanatory Test
Symbol and Standard
References
Type BF Applied Part
(Whole Device)
To identify a type BF
applied part complying
with IEC 60601-1
IEC 60417 – 5333
IEC 60601-1, Table D.1,
Symbol 20
Degree of protection against
ingress of liquid
IPX0 – not protected
against water ingress
Degree of Ingress
Protection Provided by
Enclosure
IEC 60601-1, Table D.3,
Symbol 2.
IEC 60529
Consult instructions for use
Indicates the need for
the user to consult the
instructions for use
ISO 15223 – 1. Clause 5.4.3
ISO 7000 – 1641
IEC 60601-1, Table D.1,
Symbol 11
Non-ionizing electromagnetic
radiation
The device includes a Radio
Frequency (RF) transmitter:
Make: Microchip
Module: RN42 Bluetooth FCC
ID: T9J-R942
To indicate generally
elevated, potentially
hazardous, levels of
non-ionizing radiation,
or to indicate
equipment or systems
e.g. in the medical
electrical area that
include RF transmitters
or that intentionally
apply RF
electromagnetic
energy for diagnosis or
treatment
IEC 60601-1-2 Clause 5.1.1
IEC 60417 - 5140
Direct current
To indicate on the
rating plate that the
equipment is suitable
for direct current only;
to identify relevant
terminals
IEC 60601-1. Table D.1,
Symbol 4
Rattling sound inside the unit
This is from the scrubber material
This is not an issue. Scrubber has potassium
permanganate and charcoal buds inside the device
to scrub ambient NO
FeNObreath™ Dock is
showing a red light
Indicates a fault, overcurrent,
undervoltage or overvoltage
protection circuit has triggered
Contact MGC Diagnostics Corporation or the local
distributor for assistance.