background image

PAGE 2

PAGE 3

Important Information

Before calibrating your 

FeNObreath™

, it is important to read the following information:

• 

Please ensure you charge the FeNObreath™️ for 24 hours prior to use if it has not been used in a while.

NOTE:

 The device will only charge whilst switched off.

• Please keep this Annual Service Kit at 17-24 °C (63°F to 77°F) for 24 hours 

prior to use.

• Please make sure the FeNObreath™ 

is not plugged into the mains before following any of the advice in 

this document.

• Please ensure your 

FeNObreath™

 monitor is updated with firmware version 3.0 and above. 

The latest version of firmware can be downloaded from www.nobreathfeno.com

• Please ensure the gas has not expired and that the date has not surpassed the ‘Expiry Date’ labelled on 

the CaliBag®

• Please ensure the atmospheric temperature is between 17-24 °C (63°F to 77°F)

 before calibrating. If it is 

too cold to calibrate (<17°C or 63°F)

, a blue thermometer will be shown on-screen. Move the 

FeNObreath™ 

to a warmer area and try again later. If it is too hot to calibrate (>25°C 

or 77°F), a red thermometer will be 

shown on-screen. Move the 

FeNObreath™

 to a cooler area and try again later.

Safety Information

CAUTION:

 NO scrubber contains potassium permanganate and should be disposed of as hazardous waste 

in accordance with local waste disposal regulations.

CAUTION:

 NO scrubber contains potassium permanganate and should not be tampered with and should not 

be exposed to skin.

NOTE:

 Refer to main manuals for explanations.

For any further enquiries please contact MGC by visiting: www.mgcdiagnostics.com

Attach the Bag Adapter to the rubber end of the 

CaliAdapter.

Ensure both Bag Clamps on the CaliBag® are closed 

and remove one Blue Plug.

 

Insert the Bag Adapter into the CaliBag®. 

Insert the Bung into the bottom vent hole on the 

back of the FeNObreath™️ and turn the device on.

Press the Settings icon  

 , followed by the

 

Service Area icon  

.

Press the Calibration icon  

  to begin.

Use the arrows to adjust the NO concentration

 

onscreen to match the value of the NO 

concentration written on the CaliBag®.

Press the arrow icon twice to continue.

With the CaliBag® attached, insert the CaliAdapter 

into the FeNObreath™️. Press the next arrow once 

to continue.

4

5

6

7

8

1

2

3

How to Calibrate the FeNObreath™ with the CaliBag®

Содержание FeNObreath CaliBag

Страница 1: ... to Calibrate the FeNObreath with the CaliBag Page 4 How to replace the NO Scrubber Page 6 Disposal Information Page 7 CaliBag b a c d e f g a CaliAdapter x1 b Bung x1 c CaliBag x1 d Bag Adapter e Blue Plugs x2 f Bag Clamps x2 g NO Scrubber x1 THIS KIT CONTAINS FeNObreath CaliBag ...

Страница 2: ... 25 C or 77 F a red thermometer will be shown on screen Move the FeNObreath to a cooler area and try again later Safety Information CAUTION NO scrubber contains potassium permanganate and should be disposed of as hazardous waste in accordance with local waste disposal regulations CAUTION NO scrubber contains potassium permanganate and should not be tampered with and should not be exposed to skin N...

Страница 3: ...Obreath 11 10 9 12 13 14 Please ensure the FeNObreath monitor is switched off before removing the back cover Using the screwdriver provided with the FeNObreath unscrew the screw on the back of the monitor Remove the back cover by sliding and lifting it off To remove the scrubber first press down on clip at the top left hand side of the monitor Slide out the scrubber then lift to remove Remove all ...

Страница 4: ... scrubber contains potassium permanganate and should be disposed of as hazardous waste in accordance with local waste disposal guidelines Packaging The CaliBag Annual Service Kit comes in carboard packaging this is widely recycled Please follow local waste guidelines on how to recycle or dispose of carboard The NO Scrubber comes in a plastic bag which is made of polyethylene this is also recyclabl...

Отзывы: