Important Information
Before calibrating your
FeNObreath™
, it is important to read the following information:
•
Please ensure you charge the FeNObreath™️ for 24 hours prior to use if it has not been used in a while.
NOTE:
The device will only charge whilst switched off.
• Please keep this Annual Service Kit at 17-24 °C (63°F to 77°F) for 24 hours
prior to use.
• Please make sure the FeNObreath™
is not plugged into the mains before following any of the advice in
this document.
• Please ensure your
FeNObreath™
monitor is updated with firmware version 3.0 and above.
The latest version of firmware can be downloaded from www.nobreathfeno.com
• Please ensure the gas has not expired and that the date has not surpassed the ‘Expiry Date’ labelled on
the CaliBag®
• Please ensure the atmospheric temperature is between 17-24 °C (63°F to 77°F)
before calibrating. If it is
too cold to calibrate (<17°C or 63°F)
, a blue thermometer will be shown on-screen. Move the
FeNObreath™
to a warmer area and try again later. If it is too hot to calibrate (>25°C
or 77°F), a red thermometer will be
shown on-screen. Move the
FeNObreath™
to a cooler area and try again later.
Safety Information
CAUTION:
NO scrubber contains potassium permanganate and should be disposed of as hazardous waste
in accordance with local waste disposal regulations.
CAUTION:
NO scrubber contains potassium permanganate and should not be tampered with and should not
be exposed to skin.
NOTE:
Refer to main manuals for explanations.
For any further enquiries please contact MGC by visiting: www.mgcdiagnostics.com
Attach the Bag Adapter to the rubber end of the
CaliAdapter.
Ensure both Bag Clamps on the CaliBag® are closed
and remove one Blue Plug.
Insert the Bag Adapter into the CaliBag®.
Insert the Bung into the bottom vent hole on the
back of the FeNObreath™️ and turn the device on.
Press the Settings icon
, followed by the
Service Area icon
.
Press the Calibration icon
to begin.
Use the arrows to adjust the NO concentration
onscreen to match the value of the NO
concentration written on the CaliBag®.
Press the arrow icon twice to continue.
With the CaliBag® attached, insert the CaliAdapter
into the FeNObreath™️. Press the next arrow once
to continue.
4
5
6
7
8
1
2
3
How to Calibrate the FeNObreath™ with the CaliBag®