Calibration Guide 

5

 

© Copyright 2013 coVita LLC  All Rights Reserved  v020313

 

 

30 Washington Ave Ste D Haddonfield, NJ  08033 

tel (800)707-5751 

 fax (800)721-2377 

 [email protected] 

 www.covita.net 

1
 

Firmware  

≥ V2.3

 

 

 

A.  Calibration Tips 

 

 

1.  coVita calibration instructional videos can be found 

under the support section of  

www.covita.net 

 

2.  You must calibrate every 

6 months

 to correct for 

sensor drift using 

20

 ppm 

 Carbon Monoxide

 (

CO

calibration gas.  On occasion your piCO+ may prompt 
you for an intermittent calibration. 
 

3.  Your piCO+ a has built in calibration 

reminders to prompt you to calibrate 
every 6 months. 
 

4.  Be sure to identify your 

flow meter-regulator

 for 

correct instructions. 
 

5.  Remove 

flow meter-regulator

 from can for proper 

storage and replace dust cap.  (You may hear a pop 
and this is normal.) 
 

6. 

Ensure that the caliadaptor tubing has 

no kinks

 while 

the gas is flowing.

 

B.   Required Parts 

 

 

A. 

20

 ppm 

 Carbon Monoxide

 (

CO

Calibration Can 

 

B.  Flow meter-regulator  

 

C.  Caliadaptor Tubing 

 

D.  D-Piece 

 

E.  piCO+

 Smokerlyzer

®

 

≥ V2.3

 

 

C.  Calibration Set Up

 

 

1.  Remove dust cap from can of gas 
2.  Ensure the find control valve on the flow meter-

regulator is in the off position. 

3.  Screw the regulator-flowmeter onto the gas can (some 

slight pressure may be required).  This can best done 
by screwing the gas can into the valve.  

4.  Attach the small end of the tubing onto the regulator- 

flowmeter.  

5.  Insert a D-piece into the piCO+

6.  Insert the large end of the caliadaptor tubing onto the 

D-piece. 