Hydropneumatics Subsystem
Introduction
UniCel DxC Systems Instructions For Use A13914
System Status and Commands
October 2005
Page 11-7
11
Hydropneumatics Subsystem
Introduction
Status for the Hydropneumatics Subsystem shows the status for all of the reservoirs,
canisters, and sumps. A yellow highlight indicates a warning condition and a red
highlight indicates an error condition.
The following parameters are shown:
• external DI water system status, if enabled.
• status of the various hydro float switches.
• DI water resistivity measured on the instrument and its valid range.
• actual air and vacuum pressures and their valid ranges.
The same color highlights apply.
Displaying Hydropneumatics Subsystem Status
What to Do If a Parameter is Out of Limits
The range for air and vacuum levels are only valid when the system is in Standby and
the hydropneumatics are on. The hydropneumatics are automatically turned off after
the system has been in Standby for more than 15 minutes. While the hydro is off, all
pressure and vacuum readings will show zero and a status of OK. To turn the hydro
back on, either select a prime or press the RUN button. Values will fluctuate while the
system is running.
For adjustment of the air and vacuum pressures, refer to CHAPTER 10, Maintenance,
As-Needed/As-Required Maintenance
. If the Waste B exit sump shows "Paused", it
has been stopped either automatically by the system when the Waste B container is
full, or by operator request. For additional information, refer to
Step
Action
1
Select the Status icon from the menu bar.
2
Select Hydro [F5].
3
A summary, by component name, of the monitored areas appears.