Status Summary
Introduction
System Status and Commands
UniCel DxC Systems Instructions For Use A13914
Page 11-2
October 2005
System Status
Status Summary
Introduction
Status Summary provides a high level summary of the status of the system
temperatures, power, hydropneumatic, and ICS/Smart Module components on a real-
time basis in that all parameters are updated every eight seconds. Cycle count
information and cuvette cleanliness data are also provided.
Displaying Status Summary
Possible Status Messages
The following table lists the possible status messages and their meanings. If
parameters are outside the limits, only the most severe situation will be indicated.
NOTICE
The
Status
icon on the Menu Bar is highlighted whenever a warning or error
condition exists on the Status Monitor Screen. A highlighted status icon
should be investigated by the operator, as further action may be necessary to
correct the warning or error condition.
Step
Action
1
Select the Status icon from the menu bar.
2
A summary of the monitored areas appears.
Table 11.1
Status Messages
Status
Description
Error
Parameters which exceed allowable limits are called errors. If
any of the parameters in a monitored area exceed the allowable
limit, Error will appear with a red highlight.
Warning
Parameters which are out of recommended limits but still within
allowable limits are called Warnings. If any of the parameters
in a monitored area are out of recommended limits but still
within allowable limits and there are no other parameters that
are in a error condition, Warning is shown with a yellow
highlight.
Not all parameters have both a recommended range and an
allowable range.
OK
If all of the parameters are within limits, OK is shown.