As-Needed/As-Required Maintenance
Cleaning the Work Surfaces
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-117
10
Cleaning the Work Surfaces
Clean all exposed working surfaces and covers that may have contacted sera or other
biological fluids.
5
Exit Maintenance menu.
NOTICE
To resume autoprime and return system to
Standby
, exit the
Maintenance menu.
Step
Action
, continued
(2 of 2)
Step
Action
1
Remove reaction carousel cover (1) and sample carousel cover (2).
Figure 10.78
2
Use a good grade of disinfectant soap and a soft brush, scrub covers.
3
Rinse covers in deionized water. Dry thoroughly using lintless tissue.
4
Reinstall covers.
5
Wipe all exposed surfaces on the system (including sample rack area
covers, pull down cover, compartment doors, keyboard, table surfaces and
sides of console) with a 10% bleach solution (one part sodium hypochlorite
5.25%, nine parts deionized water).
Wipe the touchscreen with 70% isopropanol, water, or any commercial
computer screen cleaner.