Three-Month Maintenance
Replace Syringe Plungers
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-49
10
Priming the Syringes
Returning to Home Position
Step
Action
1
Prime syringes 10 times by selecting:
• the Utils icon from the menu bar,
• <1> Prime,
• CC [F2],
• Reagent and/or Sample Delivery Subsystem,
• 10 in the Number of primes to repeat field,
• MC [F4],
• Sample Delivery Subsystem (10 primes are already selected).
Select <Start Prime>.
2
Check the syringes and aspiration lines for air bubbles. If bubbles persist
after priming:
Remove syringe and repeat "Installing the Syringe," Steps 4–6 and
"Priming the Syringes," Step 1.
Step
Action
1
Select the Instr Cmd icon from the menu bar.
2
Select <1> Home.
3
All instrument parts will return to their home positions.