![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 284](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768284.webp)
Two-Month Maintenance
Replace CTS Blade/Wick (CTS Option only)
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-38
October 2005
Step
Action
1
To access this procedure, select:
• the Utils icon from the menu bar
• <2> Maintenance
• <1> 1-Blade Thick CTS Blade Replacement
To start the procedure, select <Continue>.
OR
To exit this screen, select <Cancel>.
2
The instrument prepares for blade and wick replacement. Then the
Maintenance Confirmation Window comes into view.
3
CAUTION
The points of the Blade are very sharp and extend below the Wash
Tower. Stay away from the bottom of the Wash Tower.
CAUTION
DO NOT select <OK> when the cover is removed. This would move the
Cap Piercer Carriage up and cause a pinching hazard.
• Fully loosen the two captive Cap Piercer cover screws (1) at the base of
the Cap Piercer cover.
• Lift and remove the cover.
Figure 10.21
(1 of 5)
A014462P.EPS
1