Monthly Maintenance
Replace Alkaline Buffer
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-18
October 2005
Monthly Maintenance
Replace Alkaline Buffer
The following procedure describes how to replace the Alkaline Buffer reagent and the
the In-line air filters.
Step
Action
1
Loosen the two Phillips screws on the ISE module cover and remove the
cover.
2
Release the locking pin (1) on the ISE module and lift the module until it
clicks. The locking pin is located on the left side of the ISE module on
DxC 800 Systems as shown below. The locking pin is located on the right
side of the ISE module on DxC 600 Systems.
Figure 10.6
3
Unscrew and remove the bottle cap from the alkaline buffer reagent. Hold
the cap with attached straw over a container that will catch the residual
fluid from the alkaline buffer return line #33.
4
Prime the line 10 times by selecting:
• the Utils icon from the menu bar,
• <1> Prime,
• MC [F4],
• ISE CO
2
Alkaline Buffer.
Type 10 in the Number of primes to repeat field.
Select <Start Prime>. Reagent line is drained of old reagent.
(1 of 3)