Processing Parameters
First Inject: Component
User Defined Reagents
UniCel DxC Systems Instructions For Use A13914
Page 9-10
October 2005
Processing Parameters
First Inject: Component
The First Inject component designates the compartment of the cartridge which will be
pipetted first.
First Inject: Dispense Volume
The combined volume of all inject reagents must be between 200 and 327 µL. If
Second Inject and Third Inject reagents are not used, the minimum volume for First
Inject is 200 µL.
Second Inject: Component
The Second Inject component designates the compartment of the cartridge which will
be pipetted second. If Second Inject is not used, the compartment selection is None
and should not be modified.
Second Inject: Dispense Volume
The combined volume of all inject reagents must be between 200 µL and 327 µL.
Table 9.14
Options for the First Inject Component
A, B, or C
Table 9.15
Allowable Entries for the First Inject Dispense Volume
125 to 327 µL
Table 9.16
Options for the Second Inject Component
None, A, B, or C
Table 9.17
Allowable Entries for the Second Inject Dispense Volume
6 to 75 µL