Absorbance Versus Time
Introduction
Results Recall
UniCel DxC Systems Instructions For Use A13914
Page 8-12
October 2005
Absorbance Versus Time
Introduction
To get data to help troubleshoot completed chemistry results, use the Absorbance
versus Time feature.
Get Data
To get data, follow the steps below.
Step
Action
1
Recall the results of the desired sample (use any of the Results Recall
criteria described in the previous sections).
2
When the results of the desired sample come into view, select the desired
chemistry(ies). Selected chemistries are highlighted next to the chemistry
name.
3
Select ABS [F3].
Note: The Absorbance vs. Time Plot is shown for the first selected
chemistry.
4
To see the Absorbance Plot Summary Data, select the down arrow.
5
To output...
Select...
A file to a diskette
Done [F10] to exit
the screen. Then
select Disk [F1]
Absorbance vs. Time Plot and Absorbance
vs. Time Table to a printer
Both
Either Absorbance vs. Time Plot or
Absorbance vs. Time Table to a printer,
depending on which is on the screen. Use the
the Up or Down arrow, or the Plot or Table
tab at top of screen to go from the plot to the
table.
Absorbance vs. Time Table to the screen
Table tab
The plot for the next chemistry
Next
The plot for the previous chemistry
Prev