Overview
Introduction
UniCel DxC Systems Instructions For Use A13914
Results Recall
October 2005
Page 8-1
8
CHAPTER 8 Results Recall
Overview
Introduction
Results Recall allows patient and control results to be recalled, reviewed, updated, and
printed. Results can be recalled, viewed, and printed by:
• individual Sample ID or a list of Sample IDs
• rack and position
• Patient ID or
• run date and time.
Limitations
The system retains a maximum of 150,000 results and 10,000 sample programs. Once
these limits have been exceeded, the system will chronologically overwrite existing
results.
Once a sample program has been cleared, results cannot be recalled by Sample ID.
Results may still be recalled by other demographic information.
To avoid retrieving inconsistent sample data, recall results using Sample ID or verify
correct Sample ID when recalling by rack and position. Results for the last sample run
in a specific rack/cup position will be recalled. Results for the new sample will be
available for recall only after the system begins to run the new assignment.
CAUTION
If a newly assigned sample that has not been run is recalled, the results for
the previous sample run in that position will be shown or transmitted.
When using the Results Recall feature, do not leave any samples with tests
In Progress on the screen. This prevents the rack from being removed from
the system and a printout of the results from being produced.