Batch Programming
Introduction
UniCel DxC Systems Instructions For Use A13914
Sample Programming
October 2005
Page 7-9
7
Batch Programming
Introduction
Batch Programming provides programming of multiple samples with the same
chemistries, sample type, dilution factor, and status. Any Panel, Comment,
Demographics, Dilution Factor, Sample Type, or Chemistry programmed will apply to
all sample positions designated within the batch. This information may be edited at
any time on an individual position basis once the batch programming is completed.
Sample identification and description of each sample may be individually entered later
through the Sample Programming function. The Batch Mode is complete when End
Batch is requested or upon exiting the Sample Program screen.
Batch Programming
Batch programming in rack and position mode can be performed using the following
procedure.
Step
Action
1
Select the Samples icon from the menu bar.
2
Select the tests, sample type, and any other data that is common for all
samples in the batch, including sample options information.
3
Select Batch [F4]. A list of available racks for batch programming
appears. Racks may also be cleared from this screen.
To clear racks, continue with Step 4.
To batch program, continue with Step 5.
4
To clear racks, type the specific rack ID in the Rack field and select
<Clear>. A message will appear, verifying that all programming for this
rack will be deleted.
Select <OK> to delete.
OR
Select <Cancel> to keep current programming.
Either response will return the screen to the Batch programming screen.
(1 of 2)