Prior to Programming
Prior to Programming
Sample Programming
UniCel DxC Systems Instructions For Use A13914
Page 7-2
October 2005
Prior to Programming
Prior to Programming
Prior to programming, it is necessary to verify rack status and sample carousel status.
These options will determine which racks are available for programming, for rack
assignment, and for determining the instrument status.
Rack Status
Patient samples, controls, or calibrators may be assigned to racks which are then
loaded onto the system for sample processing. Each rack is uniquely numbered
(1–999) and can be assigned a maximum of four positions. The Rack Status screen
shows information about the current programming for a rack, designating how many
positions within a rack are available for programming. Racks can be viewed and
cleared from this screen. Only manually assigned samples or calibrators need to be
cleared from racks.
Step
Action
1
Select the Samples icon from the menu bar.
2
Select Racks [F8].
3
Select a rack or type a specific rack number in the Rack field to view
Sample Status. (Refer to
description of each Sample Status.) Press <Page Up> or <Page Down> to
view additional rack numbers.
4
The following functions are available:
In order to...
Action...
Clear a rack or several racks,
select Clear [F1]. Type rack or
Sample ID and select <OK>.
Select <OK> again to confirm
deletion of sample programs.
Return to Sample Programming
without making changes,
choose Cancel [F9].
Select a rack to program a
sample,
type manual assignment if
desired. Select <OK> to save.
Select a rack with an available
position and select <OK> [F10].