Calibrator Set Point Modifications
Introduction
Reagent Load/Calibration
UniCel DxC Systems Instructions For Use A13914
Page 5-30
October 2005
Calibrator Set Point Modifications
Introduction
The Calibrator Set Point Modifications option allows editing of the programmed
calibrator set points. This option can be used to alter any chemistry provided it:
• has calibration data.
• uses three or less calibrator levels.
• is configured on the system.
The user-defined chemistry set points may be modified within the user-defined
parameters setup.
Modifying Calibrator Set Points
Follow the steps below to modify calibrator set points.
NOTICE
Altering a programmed calibrator set point is not usually justified. However,
there may be a situation when a calibration value needs to be reassigned. Set
Point Modification should not be used routinely to correct instrument
correlation. The magnitude of error which is deemed acceptable when
modifying a set point is totally under the control of the laboratory and,
therefore, the Set Point Modification function should be used with caution.
Step
Action
1
Select the Rgts/Cal icon from the menu bar.
2
Select Options [F6].
3
From the Cal Options dialog box, select Modify set points using the pull-
down menu.
4
Select <Next> or <Prev> to locate the calibrator of interest.
OR
Select the Calibrator Name options button <
▼
> at the top of the screen to
view the List of Calibrators.
(1 of 2)