Reagent Load
Introduction
UniCel DxC Systems Instructions For Use A13914
Reagent Load/Calibration
October 2005
Page 5-1
5
CHAPTER 5 Reagent Load/Calibration
Reagent Load
Introduction
This chapter describes how to load reagents onto the DxC 800 and DxC 600 systems
and calibrate chemistries using the Reagent Load/Calibration Screen. The DxC system
loads and removes both cartridge (CC) and modular (MC) chemistries bulk reagents.
Reagent information encoded on the container label can be read by the bar code reader
or entered manually from the keyboard.
Cartridge reagents:
• are stored in wedge-shaped cartridges containing three separate compartments.
• include a majority of chemistries (excluding the electrolytes).
• include user-defined reagents.
Modular reagents:
• Modular reagents are stored inside the lower left compartment in either 500 mL,
1 L, or 2 L bottles.
• Other solutions associated with modular reagents include:
- No Foam (used with CREA chemistries)
- CTS Auto-Gloss (used to lubricate CTS blade)
- Wash Concentrate Buffer II
- Alkaline Buffer
No Foam, CTS Auto-Gloss, and Wash Concentrate II are stored in bottles in the center
Hydropneumatics compartment. The Alkaline Buffer is stored in a bottle within the
ISE Module.
Always check reagent status before starting a run. Make sure reagent is loaded for all
tests ordered and that all reagents have sufficient volume to complete the number of
tests in the run.
Pre-Run Checklist
Before testing samples, generate a Pre-Run Checklist to determine if reagent should be
loaded or calibrated. From the Main screen, select Pre-Run [F9] or select the Rgts/
Cal icon. Select Print [F10].
If sample programming is not downloaded to the DxC before samples are loaded,
check the LIS for pending sample programming.