![Beckman Coulter UniCel DxC Synchron 600 Скачать руководство пользователя страница 137](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768137.webp)
Auto Generation of Control
Introduction
UniCel DxC Systems Instructions For Use A13914
System Setup Options
October 2005
Page 4-45
4
Auto Generation of Control
Introduction
When this feature is enabled, if a sample is loaded with a defined control ID, the
instrument will automatically run any chemistry that is on-board and runnable for that
control.
When Auto Generation of Control is enabled, the automatic Multiple Cartridge option
is available. With the Multiple Cartridge option enabled, the system will automatically
run all cartridges on-board and runnable for any chemistries for an Auto Generated
control. Enabling Multiple Cartridge without Auto Generation will have no effect on
the manual programming of control samples.
Setup
From the Setup screen, scroll down and select <29> Auto Generation of Control to
enable/disable either feature.
The system must be in Standby or Stopped to make modifications.
The default is disabled.