A13914AF
11-13
Utilities
Backup/Restore
11
Backup/Restore
Introduction
The Backup/Restore option provides the ability to store data to a diskette for use at a later time,
should an error occur that causes the loss of data. Backup is the process of saving data to a diskette
for safe storage. Restore takes previously backed up data from a diskette and places it back into the
system. The information is divided into two categories: System Parameters and Alignment Data,
each stored on a separate diskette.
System Parameters
System Parameters contains setup, e-maintenance log, reagent, and calibration information.
Alignment Data
Alignment Data includes all alignment information including photometer data.
Backup the System
The backup procedure should be performed following installation of a new software version, a
change in alignments or a change in system parameters, especially setup parameters. To perform
a backup, follow the steps below.
NOTE
Alignment files are on a separate diskette from System Parameters. Depending on the instrument
configuration, System Parameters may require more than one diskette.
IMPORTANT
Do not start a backup of System Parameters or Alignments while the printer is printing. If the
printer is printing, it will correctly stop before backup but will not restart when the backup is completed.
A reboot will be required to return the system to normal operation.
1
Select:
•
Utils
from the menu bar,
•
8 Backup/Restore
,
OR
Type
8
in the
Option Number
field and press
(
Enter
)
.
2
Select the
System Parameters
box.
AND/OR
Select the
Alignment Data Files
box.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......