A13914AF
9-75
Maintenance
As-Needed/As-Required Maintenance
9
5
Locate the CO
2
measuring electrode. Place absorbent tissue beneath the CO
2
electrode port.
Using hemostats, clamp off reagent Lines #33 (1) and #32 (2) as closely as possible to the CO
2
electrode. (Line #32 connects as a loop to the CO
2
reference electrode.) Remove electrode by
turning the electrode retainer nut.
Figure 9.53
6
Remove electrode retainer nut (1). Remove membrane as follows:
•
Separate the membrane retainer (3) assembly from the electrode.
•
Remove the quad-ring (4) from the retainer. Inspect the electrode port if the quad-ring is
not on the electrode.
•
Separate the membrane retainer from the membrane clamp (2) and pull apart. Discard the
used membrane.
Figure 9.54