A13914AF
9-60
Maintenance
Six-Month Maintenance
8
Set the glucose sensor down into an empty area of the cup modules. Allow the free standing
sensor to remain connected but not installed for 24 hours. Be careful that the sensor cable is not
pinched or crimped. Reinstall the MC module and ISE module covers.
9
Proceed to Draining Reaction Cup for Sensor Maintenance or Replacement below to replace the
sensor.
Drain the Reaction Cup for Sensor Maintenance or Replacement
IMPORTANT
System does not autoprime in
Maintenance
mode. After completing the maintenance
procedure, or if a procedure is aborted, exit the Maintenance menu to resume autoprime and return the
system to
Standby
.
1
Drain the reagent from the MC cup.
•
Select:
—
Utils
from the menu bar,
—
2 Maintenance
,
—
4 Cup Maintenance
.
2
Select
Glucose
cup to be drained in preparation for maintenance.
3
Select
Drain
. This drains the cup of reagent so that maintenance may be performed,
OR
Select
Cancel
to immediately return to the Maintenance menu without draining the cup.
4
Proceed to Sensor Replacement and Stir Bar Cleaning.
Sensor Replacement and Stir Bar Cleaning
Follow the steps below to replace the sensor. After replacement, perform a Sensor Calibration, then
calibrate Glucose.
IMPORTANT
When a new sensor is installed, if needed, the reaction cup and/or reagent lines can be cleaned
at this time. Follow the Four-Month Maintenance procedures found in the Four-Month section of this
chapter.
1
Remove the ISE module and MC module covers.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......