A13914AF
9-59
Maintenance
Six-Month Maintenance
9
IMPORTANT
A new glucose sensor may drift over the first 24 to 48 hours. Always verify control recoveries
after calibration. More Frequent calibration may be required during the first 24 to 48 hours following
installation.
CAUTION
Do not touch membrane tip.
1
Remove the AccuSense glucose oxygen sensor (PN 467432) from the package.
2
Remove the protective cap from the electrode connector.
3
Check the sensor for the presence of the gasket on the membrane tip. If the gasket is missing,
replace it at this time.
4
Visually inspect the sensor for leakage or puncture of membrane.
5
Hold the sensor in a vertical position with the membrane tip pointed down. Tap on the side of
the sensor body to remove any bubbles from the membrane tip.
6
Remove the ISE module and MC module covers.
7
Push the glucose sensor connector into the empty connector port (1) of the BUNm/UREAm
module until the latch clicks.
Figure 9.40
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......