Maintenance
Six-Month Maintenance
ISE Service
The ISE service procedure prevents reagent leakage from the flow cell whenever electrode removal
is required.
1
Loosen the two Phillips screws (1) on the ISE module cover and remove the cover.
Figure 9.34
2
Release the locking pin (1) on the ISE module and lift the module until it clicks into a raised,
locked position. The locking pin is located on the left side of the ISE module on DxC 800 Systems
as shown below. The locking pin is located on the right side of the ISE module on DxC 600
Systems.
Figure 9.35