Maintenance
Monthly Maintenance
9
Place a fresh bottle of alkaline buffer on the system. Wipe reagent straw and replace bottle cap
onto the new reagent bottle.
Figure 9.9
10
Load reagent information from bottle bar code into the system either automatically using the
hand-held bar code reader, or manually using the keyboard. Refer to
11
Select
Done F10
when finished.
12
While priming with the alkaline buffer reagent, check the peri-pump tubing and line for leaks
or other problems. Correct as needed.
13
Adjust the fluid level of the Alkaline Buffer Damper Assembly. Refer to the
of the Alkaline Buffer Damper Assembly
procedure.
CAUTION
Pinch hazard. Keep fingers clear of the ISE module as you lower it.
14
Lower the ISE module by releasing the locking pin. Be careful NOT to pinch tubing as the module
is lowered.
15
Reinstall the ISE module cover.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......