A13914AF
8-19
User Defined Reagents
User Defined Reagent Removal
8
•
If a single trigger is defined, it should be defined in the third inject column. If it is defined in
the second inject column, the system transfers the information to the third inject column upon
exit.
User Defined Reagent Removal
Remove a User Defined Reagent
1
Remove UDR reagent cartridge from system.
2
Delete UDR information from the following areas:
•
reference ranges
•
calibration assignments
•
sample programming and
•
quality control.
3
Select
Setup
from the menu bar.
4
Select
2 Chem Config/Sample Type
.
5
Select the chemistry to be deleted.
6
Select
Delete F6
to remove the UDR from the chemistry menu.
7
Select
Setup
from the menu bar.
8
Select
15 User-Defined Chemistries
.
9
Select the
Option No.
of the chemistry to be deleted.
10
Select
Delete F2
.
A user defined reagent must be deconfigured from the chemistry menu prior to removal.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......