A13914AF
7-1
CHAPTER 7
Results Recall
Overview
Introduction
Results Recall allows patient and control results to be recalled, reviewed, updated, and printed.
Results can be recalled, viewed, and printed by:
•
individual Sample ID or a list of Sample IDs
•
rack and position
•
Patient ID or
•
run date and time.
Limitations
The system keeps a maximum of 150,000 results and 10,000 sample programs. When these limits are
exceeded, the system chronologically overwrites existing results.
When a sample program is cleared, results cannot be recalled by Sample ID. Results may still be
recalled by other demographic information.
To avoid retrieving inconsistent sample data, recall results using Sample ID or verify correct Sample
ID when recalling by rack and position. Results for the last sample run in a specific rack/cup
position will be recalled. Results for the new sample is available for recall only after the system
begins to run the new assignment.
CAUTION
If a newly assigned sample that has not been run is recalled, the Results for the
previous sample run in that position is shown or transmitted.
When using the Results Recall feature, do not leave any samples with tests In
Progress on the screen. This prevents the rack from being removed from the
system and a printout of the Results from being produced.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......