Quality Control
QC Chart (Levey-Jennings)
5
5
The chemistries defined for the selected control are listed. All chemistries appear to be
selected.
For convenience, there are two ways to select the chemistries to chart:
•
To chart most of the chemistries shown, select all chemistries which should NOT be
charted. Then select
Remove
.
•
To chart less than half of the chemistries shown:
—
Select
Remove All
.
—
Select the chemistries to chart.
Multiple selections are allowed. Use the scroll bar to access additional chemistries.
NOTE
User-defined chemistries which have been cleared cannot be retrieved for QC Chart.
•
Select
OK
to show the QC Chart(s).
•
Select
Cancel
to cancel the Chart request.
6
Select
Next F8
to access all requested QC charts. Each of the requested charts appear in the
order of the control chemistry configuration.
7
Select
Print F9
to print a copy of the Chart.
8
Select
Chems F1
to specify additional chemistries for charting.
9
Select
Filenum F2
to specify additional file numbers for charting.
10
Select
Done F10
to exit the
QC Chart
dialog box.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......