A13914AF
4-27
Reagent Load/Calibration
Calibration Acceptance Limits
4
Calibration Acceptance Limits
Introduction
Calibration absorbance or ADC values are compared to preprogrammed back-to-back, span, and
range limits to determine acceptable performance. If calibration errors in accuracy, precision,
sensitivity, or linearity are detected by the system computer, error flags are generated. Calibration
Acceptance provides a reference to the acceptable limits for calibration.
View Calibrator Acceptance Limits
1
Select
Rgts/Cal
from the menu bar.
2
Select
Options F6
.
3
Select
Calibrator acceptance limits
from the pull-down menu.
Chemistries appear with chemistry name, setpoint value(s), and appropriate limits.
4
Select
Next
or
Prev
to locate the calibrator of interest.
OR
Select the
Calibrator Name
pull-down menu at the top of the screen to view the List of
Calibrators.
5
Select
Close
to exit the screen.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......