A13914AF
4-14
Reagent Load/Calibration
Calibration Status
1
Select
Rgts/Cal
from the menu bar.
2
Select
Assign F7
.
3
Use the down arrow option button to open list of calibrators.
4
Select the calibrator.
5
Enter the ID of the calibrator or the rack/position number.
•
If using a bar code label, the label must match the assigned calibrator ID.
•
If a rack/position is assigned, the calibrator must be placed in this assigned position.
6
Repeat Steps 3–5 to assign additional calibrators.
7
When calibration assignment is complete, select
Close
.
Calibration Status
Introduction
Calibration status, as well as reagent status, is accessed from the
Rgts/Cal
dialog box. Calibration
status is shown to determine which chemistries require calibration.
Calibration Status Messages
Table 4.2
Calibration Status Messages
Message
Explanation
Requested
Indicates that calibration is pending and calibrators have been identified.
Calibrated
Indicates that the chemistry has been calibrated and has not exceeded its
calibration time period.
Cal Required
Appears when:
•
a new cartridge is loaded.
•
a user-defined cartridge is loaded.
•
a non-overrideable calibration failure has occurred.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......