4-7
Troubleshooting and Maintenance
Figure 4-5. Location of Diagnostic Messages on the Control Panel
CLEANING
Keep instrument surfaces clean by wiping them with a cloth damp-
ened with a mild detergent such as Beckman Solution 555™ (diluted
10 to 1 with water). If salts or other corrosive materials are used, or if
spillage occurs, wash all affected areas immediately. Do not allow
corrosive materials to dry on the instrument. (Be careful not to spill
liquid on the instrument where electrical or mechanical components
could get damaged.)
ROTOR CHAMBER
The rotor chamber is coated with epoxy resin paint, and the chamber
door O-ring is Buna N rubber. The chamber can be wiped clean with
a cloth dampened with a mild detergent such as Solution 555. The
O-ring should be cleaned with a tissue or soft cloth every 3 or 4
months. Replace the O-ring whenever it becomes worn or damaged.
Lightly coat the O-ring with silicone vacuum grease (335148) before
reinstalling it to ensure an optimum vacuum seal. (Instrument O-rings
have not been designed as bioseals for aerosol containment.)