773232AH
10-48
Using Navios Software
File Menu
Show % in Region on Plots
When the show % in region option is selected, the percentage of events in a particular region is
automatically displayed on new plots. For new plots all regions created will display the % gated
value of that region. When this option is selected, the % is also displayed on all existing plots in the
protocol.
Black Plot Backgrounds
This displays the plots with a black background instead of white. This can be useful as it makes the
viewing of smaller populations of cells much easier. When you print, drag or copy a plot to other
software programs, the background remains white.
Opaque Region Labels
This option displays an area around the annotation within black backgrounds, making it visible.
Print Plot Frames
This option can be used to show or hide frames on plots printed directly from the File Print menu
and exported plot images.
If this is checked, all frames are displayed around plots. If it is clear, frames are not printed or
exported.
Print Dots Black
Select this option to print all dots in a color dot plot in black irrespective of their colors. If it is clear,
dots are printed in various shades of gray or, if using a color Printer, the best matched colors
available. This option does not affect the printing of FlowPAGEs.
Log Scale Density Plots
Selecting this option displays a density plot with log scaling of the density levels. If unchecked,
linear scaling of the density levels is used. These changes are only applied to plots created after this
option has been selected, and not to the ones previously created.
Smooth Histograms
This automatically smooths single histograms by default allowing for easier viewing. When
selected, this option displays smoothed data on all new histograms as they are created. It does not
apply any of the changes to previously created plots.
Default % of Events on Dot Plot
This allows you to set the default number of dots to be displayed when acquiring data in a dot plot.
These changes are only applied to plots created after this option has been selected, and not to the
ones previously created. Reducing the default percentage increases the speed of software functions
when reading large listmode files.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......