3-21
Installing and Calibrating an Aperture Tube
The Change Aperture Tube Wizard
3
Saving the Calibration File
After you have calibrated the aperture (page
), the calibration results appear in the Viewing
Area. Results are displayed as a size graph.
To save the calibration run:
1.
Select
RunFile > Save
on the Run Menu bar, or click
(
Save
)
in the Main Toolbar.
2.
In the Save As window, navigate to the appropriate folder.
3.
In the File name field, enter a name for the calibration file.
NOTE
The Multisizer 4 software uses the .#m4 extension for all run data files, including calibration
runs.
4.
Click
(
Save
)
.
The Change Aperture Tube Wizard
The Change Aperture Tube Wizard in the Multisizer 4 software provides step-by-step instructions
to guide you through the steps you need to follow when installing a new aperture tube. The Wizard
allows you to verify aperture calibration, test aperture settings, and enter the new aperture tube
size and serial number.
Use the Change Aperture Tube Wizard after you have:
•
Powered ON the Analyzer and connected the software to the Analyzer
•
Selected the aperture tube size
•
Calibrated the aperture, and
•
Emptied the Waste Tank
Содержание Multisizer 4
Страница 12: ...xii List of Figures ...
Страница 14: ...xiv List of Tables ...
Страница 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Страница 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Страница 90: ...PN A51387AB 3 36 Installing and Calibrating an Aperture Tube Running a Concentration Control Sample ...
Страница 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Страница 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...