Installing and Calibrating an Aperture Tube
Installing an Aperture Tube
Installing an Aperture Tube
This section describes how to correctly place an aperture tube in the Analyzer. After installing the
aperture tube, you will need to:
•
Calibrate the aperture
•
Follow the step-by-step guide provided in the Change Aperture Tube Wizard. The Change
Aperture Tube Wizard allows you to verify calibration, test aperture settings, and enter the new
aperture tube size and serial number in the Multisizer 4 software.
To install an aperture tube:
1.
Clean the new ST Aperture Tube. For instructions on the proper aperture cleaning procedure,
see
Cleaning the Aperture Tube, Beakers, and Accuvette® STs
.
Table 3.1
Aperture Sizes
Aperture
(Nominal
Diameter,
µm)
Diameter Range (µm)
Volume Range µm
3
or fL
Total
Standard
Extended
Total
Standard
Extended
20*
0.4–16
0.4–12
12–16
0.034–2.14
×
10
3
0.034–905
905–2.14
×
10
3
30*
0.6–24
0.6–18
18–24
0.113–7.24
×
10
3
0.113–3.05
×
10
3
3.05
×
10
3
–7.24
×
10
3
50
1.0–40
1.0–30
30–40
0.524–33.5
×
10
3
0.524–14.1
×
10
3
14.1
×
10
3
–33.5
×
10
3
70
1.4–56
1.4–42
42–56
1.44–92.0
×
10
3
1.44–38.8
×
10
3
38.8
×
10
3
–92.0
×
10
3
100
2.0–80
2.0–60
60–80
4.19–268
×
10
3
4.19–113
×
10
3
113
×
10
3
–268
×
10
3
140
2.8–112
2.8–84
84–112
11.5–736
×
10
3
11.5–310
×
10
3
310
×
10
3
–736
×
10
3
200
4.0–160
4.0–120
120–160
33.5–2.14
×
10
6
33.5–905
×
10
3
905
×
10
3
–2.14
×
10
6
280
5.6–224
5.6–168
168–224
92.0–5.88
×
10
6
92.0–2.48
×
10
6
2.48
×
10
3
–5.88
×
10
6
400
8.0–320
8.0–240
240–320
268–17.2
×
10
6
268–7.24
×
10
6
7.24
×
10
6
–17.2
×
10
6
560
11.2–448
11.2–336
336–448
736–47.1
×
10
6
736–19.9
×
10
6
19.9
×
10
6
–47.1
×
10
6
1000**
20–800
20–600
600–800
4189–268
×
10
6
4189–113
×
10
6
113.1
×
10
6
–268
×
10
6
2000**
200–1600
200–1200
1200–1600
4.19
×
10
6
–2.14
×
10
9
4.19
×
10
6
–905
×
10
6
905
×
10
6
–2.14
×
10
9
* Range depends on system cleanliness and environmental electrical noise.
** Range depends on sample density.
Содержание Multisizer 4
Страница 12: ...xii List of Figures ...
Страница 14: ...xiv List of Tables ...
Страница 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Страница 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Страница 90: ...PN A51387AB 3 36 Installing and Calibrating an Aperture Tube Running a Concentration Control Sample ...
Страница 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Страница 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...