PN A51387A
B
3-1
CHAPTER 3
Installing and Calibrating an Aperture Tube
Overview
When to Install a New Aperture Tube
Install a new aperture tube if:
•
You are using the Analyzer for the first time
•
You are using different aperture tube sizes to obtain different particle size distributions
•
Your aperture tube has broken
To install a new aperture tube:
1.
Select an aperture tube size
2.
Use the Change Aperture Tube Wizard to install the new aperture tube
Selecting an Aperture Size
Select an aperture tube size based on the size of the particles you will be measuring. The Analyzer
can accurately measure particles between 2% and 80% of the aperture size, and in the range of 10%
to 80% of the aperture diameter for a 2000-µm aperture tube.
CAUTION
If you select a 20-µm or 30-µm aperture, Beckman Coulter recommends that you
follow special cleaning, handling, and operating procedures to accommodate the
sensitivity of small apertures. For more information, see
, page
.