Analyzer Overview
Cleaning the Analyzer
Figure 1.9
Cleaning an Aperture Tube
Cleaning the Analyzer
Clean the system thoroughly before running analyses. An effective cleaning procedure includes
filling and draining the system multiple times and in multiple cycles, alternating between de-
ionized water and 98% Isopropyl Alcohol (IPA).
Drain the system:
1.
On the Multisizer 4 software Main Menu bar, select
Run > Drain System
.
2.
Remove and empty the Electrolyte and Waste Jars.
3.
Replace the Waste Jar.
Clean the system with de-ionized water:
4.
Rinse and fill the Electrolyte Jar with clean or pre-filtered de-ionized water.
5.
Replace the Electrolyte Jar.
6.
On the Multisizer 4 software Instrument Toolbar, click
(
Empty
)
to empty the Waste Tank.
7.
On the Multisizer 4 Main Menu bar:
a.
Select
Run > Fill System
, or click the
(
Fill
)
button on the Instrument Toolbar.
b.
When the system has filled, select
Run > Drain System
.
c.
Fill and drain the system two more times (for a total of three times).
8.
Remove and empty the Electrolyte Jar.
Содержание Multisizer 4
Страница 12: ...xii List of Figures ...
Страница 14: ...xiv List of Tables ...
Страница 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Страница 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Страница 90: ...PN A51387AB 3 36 Installing and Calibrating an Aperture Tube Running a Concentration Control Sample ...
Страница 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Страница 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...