PN 4277248DD
4-18
SPECIFICATIONS/CHARACTERISTICS
BAR-CODE LABEL SPECIFICATIONS
4.6
BAR-CODE LABEL SPECIFICATIONS
General
The LH 700 Series System supports the use of of up to 16-character bar-code labels. A
bar-code consists of black lines (bars) and white lines (spaces), which are called elements.
There are narrow elements (NE) and wide elements (WE); their arrangement is determined
by the code.
Optical Characteristics at 880 nm ±10% and 633 nm ±10%
r
Print Contrast Signal (PCS): 80% min.
r
Reflectivity of Media (RW): 80% min.
r
Reflectivity of Ink (Rb): 16% max.
r
No spots or voids; no ink smearing.
r
Edge roughness is included in the bar and space tolerances.
Measurement method is according to American National Standards Institute's
MH10-8M-1983.
Printing Method
Photographic, or thermal transfer.
Label Thickness
Maximum label thickness must be such that:
r
The tube's outer diameter including the label is not greater than 13.3 mm.
r
The label including adhesive = 0.006 ±0.003 in.
NE/WE Ratio
Must remain constant over code length.
IMPORTANT
1. DO NOT use the following characters # @ [ \ ]
`
{ | } ~ in Specimen or Patient identifiers.
There is a potential for Specimen or Patient misidentification to occur. The system will substitute or omit
these characters when the system in configured in a language other than English or Chinese.
2. Risk of misidentification. Use of poor quality, dirty, improperly placed or damaged bar-code labels could
keep the instrument from reading the bar-code labels. Ensure the bar-code labels are undamaged. Ensure
the bar-code labels conform to the specifications provided in Chapter 4 of the Reference manual.
3. DO NOT use leading or trailing spaces in the ID.
IMPORTANT
Possible misidentified results. For accurate reading by the scanner, it is important that
bar-code labels for specimen tubes adhere strictly to the specifications given in this section. Labels that
meet these specifications are available from Beckman Coulter.
Содержание COULTER LH 750 System
Страница 26: ...PN 4277248DD 1 12 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Страница 54: ...PN 4277248DD 3 22 OPERATION PRINCIPLES XB ANALYSIS ...
Страница 84: ...PN 4277248DD 4 30 SPECIFICATIONS CHARACTERISTICS BAR CODE LABEL SPECIFICATIONS ...
Страница 90: ...PN 4277248DD 5 6 HAZARDS DISPOSAL OF ELECTRICAL INSTRUMENTATION ...
Страница 116: ...PN 4277248DD 6 INDEX INDEX X XB Analysis definition GLOSSARY 16 XM Analysis definition GLOSSARY 16 ...
Страница 118: ......