INSTRUCTION MANUAL 

 

HIGH PRESSURE PNEUMATIC PUMP 

MODEL PG

PH 

 

 

 

 

Содержание PGPH

Страница 1: ...INSTRUCTION MANUAL HIGH PRESSURE PNEUMATIC PUMP MODEL PGPH ...

Страница 2: ......

Страница 3: ...ty of the contents of this manual For more detailed technical data about the Instruction manual for Beamex PGPH High Pressure Pneumatic Pump please contact the manufacturer Copyright 2013 2015 BEAMEX OY AB Ristisuonraitti 10 FIN 68600 Pietarsaari FINLAND Tel 358 10 5505000 Fax 358 10 5505404 E mail sales beamex com service beamex com Internet http www beamex com 8804400 PGPHuEng Version 1 1 ...

Страница 4: ......

Страница 5: ...ram 3 OPERATION 3 Keep in Mind 4 TROUBLESHOOTING MAINTENANCE 5 Seal Replacement 6 Cleaning the Pump 6 Leak Tests 6 Leak Test for Positive Pressure 6 Leak Test for Negative Pressure Vacuum 6 SPECIFICATIONS 7 The Pump Unit 7 Pressure Measurement Hoses Optional Accessories 8 Connections 8 WARNINGS AND CAUTIONS 9 ...

Страница 6: ...connect ing the instru ment to be cali brated to PGPH Connections when calibrating a transmitter using an internal pressure module Communication cable Hose connector set for connect ing the instru ment to be cali brated to PGPH Connections when calibrating a transmitter using an external pressure module ...

Страница 7: ... connectors A set of seals o rings type NBR70 for Pressure connectors size 6 2 Guarantee leaflet This Instruction Manual Optional Accessories The optional accessories are as follows ITEM ORDER CODE Carrying case 8003315 Service seal kit 8003180 Tube of grease for the Fine Adjust stem 0005180 400 bar hose connector set for connecting the instru ment to be calibrated to PGPH 8003380 400 bar hose con...

Страница 8: ...ure connector G for instrument calibrator Hand tightened Vent connection hole Fine Adjust handwheel Rotating clockwise increases pressure Pressure connector G for instrument calibrator Hand tightened Equivalent with Pressure connector Pressure Vacuum selector Pull out for pressure and push in for vacuum Do not change when pressurized Selector lock Pump body ...

Страница 9: ...s and are equivalent See also Leak Tests on page 6 Calibration 6 Pump pressure vacuum 7 Close Shut off valve clockwise See also notes further on 8 Adjust pressure by rotating the Fine Adjust handwheel 9 Repeat stages 6 to 8 until instrument s maximum pressure deepest vacuum is reached When maximum pressure deepest vacuum is reached continue as follows 10 Gently open Release valve to release pressu...

Страница 10: ...e pump lever but no pressure change is indicated stop pumping and locate the fault Always keep a reliable indi cator connected to the measurement system If the pump fails to indicate a pressure change after considerable pumping action of the handle check to assure that the connections between the pump the hose end and the attached instrument s are tight and retry pumping Also check to assure that ...

Страница 11: ...Check Pressure connectors and Re tighten if necessary Hand tightening should be sufficient The o ring in Pressure connectors or is missing misplaced or broken Replace the o ring The end surface of the instrument s con nection thread is not smooth Use a PTFE washer in Pressure connector The instrument s calibrator s or hose s thread does not match the thread on PGPH s Pressure connectors or Use pro...

Страница 12: ...le to the pump Plug any open connections You ll also need a clock to measure time Leak Test for Positive Pressure Increase pressure using constant change rate up to pump s full scale positive pres sure 140 bar 2000 psi Start your clock After five minutes record the pressure indication p5 After ten minutes record the pressure indication p10 Calculate the leak Lp using this equation 5 10 5 p p p p L...

Страница 13: ...elief Valve PGPH includes a built in relief valve Pressure media Clean air Temperature 0 50 C Humidity 85 RH Pressure resolution 0 1 mbar 0 002 psi Output connectors Two G female connectors One for the instrument to be calibrated and another for the calibrator Clean out connection G female plugged If local atmosphere pressure is 1 bar 14 5 psi the vacuum can reach to 0 95 bar 14psi If local atmosp...

Страница 14: ...ar 195 0 MPa 28200 psi Operable temperature range 20 C to 100 C 4 F to 212 F Smallest bending radius 20 mm 0 78 below 20 C 4 F 30 mm 1 18 Connections End connectors in 400 bar hose connector set for connecting the instrument to be calibrated to PGPH Code 8003380 PGPH end G1 4 male Instrument end G 1 4 NPT male End connectors in 400 bar hose connector set for connecting a calibrator s high pres sur...

Страница 15: ...b serve the effect of the operational conditions to the maximum pressure allowed in the hose The specifications of Beamex hoses can be found on page 7 of this manual The environmental conditions may restrict the allowable maximum pressure to a lower level than the pump and the hose enable In that case please consider using a relief valve Always depressurize PGPH when it is left on its own Avoid po...

Страница 16: ...ttp www beamex com Beamex Inc 2152 NW Parkway Suite A Marietta GA 30067 U S A Phone 800 888 9892 1 770 951 1927 Fax 1 770 951 1928 E mail beamex inc beamex com Beamex Limited Newtown Grange Farm Business Park Desford Road NEWTOWN UNTHANK Leicestershire LE9 9FL United Kingdom Phone 01455 821 920 Fax 01455 821 923 E mail beamex ltd beamex com Representative ...