background image

Quick Set-Up Guide: TotalAlert Infinity

 Master Alarm

4107 9016 92.01

1

Enter

Backspace

Close

Clear all 

entered 

text

Type the default password (121212) and touch 

Enter. Choose continue or change password.

2

3

4

6

8

9

11

12

13

15

16

Select number of source locations for the above se-

lected gas (1 location per piece of source equipment)

To select alarm description text:

• 

Scroll through the options using the right and left 

arrows at the bottom of the screen. 

• 

Touch the tab for the appropriate option.

• 

Touch the Save button when you see a check beside 

your selection to close and return to the SIGNAL 

SETTINGS screen. 

• 

Touch the X button in the upper right to save the 

signal text and return to the CONFIGURE (SOURCE) 

screen to set up remaining signals for the first 

location.

• 

Set up remaining locations for the first gas (badge 

1), then set up additional gases as needed following 

steps 6-15.

• 

To return to the home screen, touch the heartbeat 

icon.

7

Clear all 

entered text

Cancel

After inputing source location, touch save, then select 

the location to turn it green. Touch save.

14

10

Heartbeat 

Icon

Alarm

Information History

Configuration

5

Save

Clear all 

entered text

Cancel

Enter master alarm location and touch save.

Select appropriate gas and color code and touch save.

Route each pair of signal input 

wires as shown to appropriate 

terminals on input board(s). 

Touch the corresponding signal 

point and touch save once the 

point has been selected.

Note: Alarm may sound. Touch 

the mute button to silence the 

alarm.

Mute Button

Touch B to set the LOCATION text.

Touch 1 to set the gas and source locations, which will 

be shown on badge 1.

Touch a blank location label to enter new text, or choose 

an existing location label if the location has already been 

set up. Touch save.

Touch the B1 tab to begin set up of the first alarm point 

(B). 