User manual

Version 5.2

Bioenergy TUMBLER 1000+

Pellets durability tester acc. EN ISO 17831-1

Содержание TUMBLER 1000

Страница 1: ...User manual Version 5 2 Bioenergy TUMBLER 1000 Pellets durability tester acc EN ISO 17831 1 ...

Страница 2: ... UNPACKING 6 3 2 INSTALLATION 6 3 3 FIRST OPERATION 6 4 OPERATING INSTRUCTIONS 7 4 1 GUIDELINES FOR CLEANING 9 5 GUIDELINES FOR TESTING 10 6 PRINCIPLE 10 7 CALIBRATION CHECKING 10 8 FAULT FINDING 11 9 TECHNICAL SPECIFICATIONS 12 10 WIRING DIAGRAM 13 11 DECLARATION OF CONFORMITY 14 12 BILL OF MATERIALS 15 13 CONTACTS 16 ...

Страница 3: ...User manual TUMBLER 1000 2 1 Pictorial description 1 Main switch ON OFF 2 START STOP Buttons 3 Emergency stop switch 4 Indicator light 5 Apparatus door 6 Level adjusting screw 7 Display 1 2 3 4 5 6 7 ...

Страница 4: ...User manual TUMBLER 1000 3 1 Tumblerbox 2 Locking latch adjustable 3 Door contact switch 4 Switch box 5 Dust bin 5 3 3 2 4 1 ...

Страница 5: ...User manual TUMBLER 1000 4 1 Specification plate 2 Main body 3 Switch box 4 Stand 5 Air vent grid 6 Drive vent grid 7 Power plug incl fuse 8 Plate for electronic sensor inside switch box 8 1 4 2 5 3 6 7 ...

Страница 6: ...ept away from the connected apparatus The working area has to be secured in an appropriate way and must meet the local operating safety regulations Do not operate the apparatus in wet or damp atmosphere Do not operate the apparatus without the assembled dust bin Ear and eye protection is recommended during any testing Do not open the apparatus door during operation Before any cleaning or servicing...

Страница 7: ...ed with the locking latch The locking latch can be adjusted by screwing if necessary Ensure that the dust bin is placed under the tumblerbox and the apparatus door is closed properly Switch on the On Off switch and pull out the emergency stop switch until the click yellow ring is visible The tumbler starts an adjusting sequence and stops automatically afterwards The indicator light will be on and ...

Страница 8: ...the front As soon as the door is open the display shows Door Open If the tumblerbox is not in home position so that you are unable to open the locking latch or take out the box close the apparatus door and switch off the main switch and on again The tumbler starts an adjusting sequence and stops automatically in home position so that the box can be taken out Fill the tumblerbox with 500 10g pellet...

Страница 9: ...and weighed The difference between initial weight and output weight is the abrasion If the measurement is stopped by the Stop button the display first shows stopping and the stopped Do not open the apparatus door before the tumbler fully stopped After opening the door the display switches to the initial position and will again show the amount of turns and time remaining before starting the test To...

Страница 10: ...periodically according to the frequency of usage Therefore you should take out the tumblerbox first for easier handling On the back side of the apparatus is a small blue lever on the air vent grid see page 4 pos 5 which has to be switched up to be able to change the filter The filter has to be controlled periodically respectively changed as needed ...

Страница 11: ...ent is independent of any fluctuation in power frequency The tumbler will always turn 500 rotations and 10 minutes and stop automatically afterwards However a periodic check e g once a year of the specified functions is recommended Therefore you have to open the switch box with the switch box key Then the tumbler has to be started With an appropriate stop watch it has to be checked that the tumble...

Страница 12: ...Apparatus door opened too early Push the start button the tumbler turns with creep speed into home position Extreme dust formation Cleaning interval too long Defect sealings Clean the tumbler in shorter intervals Blow out exhaust the dust bin Change sealings Tumbler does not stop automatically Defect control Contact the service technician Tumbler does not stop within the specified time shorter or ...

Страница 13: ...een 7035 light grey Sound level 78 dB A measured with 500g wood pellets 6mm Power supply Voltage 240 V Frequency 50 Hz Power 180 W Fuse rating 1 5 A Note It is recommended to wear ear protection Dimensions of the stressed pellets The maximum dimensions of the pellets tested within the tumbler are diameter 12 mm length 6 cm maximum load in the box 750g ...

Страница 14: ...User manual TUMBLER 1000 13 10Wiring Diagram ...

Страница 15: ...User manual TUMBLER 1000 14 11Declaration of Conformity ...

Страница 16: ...14 RS Components Ges m b H CCCN 85366990 489122 Isolierhülle IEC Gerätestecker RS Components Ges m b H CCCN 85472000 801932 Feinsicherung 5x20 RS Components Ges m b H 1 6A träge 563728 Kompaktschaltschrank Rittal GmbH AE 1045 RAL7035 AE 1045500 Austrittsfilter Rittal GmbH SK 3237200 Gebotschild Netzstecker ziehen Kroschke Ges m b H N 23088 Spannverschluß rostfrei Peter Ofner GmbH 2 41 02 30 6 Rast...

Страница 17: ...ogen 96 1090 Vienna Austria Tel 43 676 61 261 73 Fax 43 1 4083451 E Mail markus bergler gmail com Sales Service BEA Institut für Bioenergie GmbH Avedikstrasse 21 1150 Vienna Austria Tel 43 1 890 93 91 Fax 43 1 890 93 92 E Mail office bioenergy co at Web www bioenergy co at www bioenergy institute ...