Processing Conventional Slides on the BD FocalPoint™ Slide Profiler
175
4
Use the scroll bar, to scroll down and view that statistics for each operating mode.
5
If the
Process
Review
rate value is greater than 10%, print a Process Review report for the
month. Look over the report for trends to determine if there is something to do to correct the
problem.
6
If analysis of the Process Review report doesn’t suggest any corrective actions, Contact
BD Technical Support. BD may examine the System Logs to determine the cause of the
process failures and propose corrective action to reduce the frequency of slide rejections.
18.9 Report Options
This section discusses the various types of reports and the rationale for selecting certain types of
reports and options. Because each laboratory handles their cytology work flow differently, it is
impossible to standardize a single report that meets each lab’s needs. Thus, we’ve provided
several types of reports with sorting and layout options that allow printing slide result reports that
best suit the workflow in your laboratory.
18.9.1
Report Types
There are six types of report available:
•
Slide Result Report:
Lists results for all slides. (See Figure 12
3 for an example of this
report).
•
Review Report:
Lists all slides designated
Review
.
(See Figure 12
5 for an example of this
report).
•
QC Review Report:
Lists all slides selected for
QC Review
plus a limited number of
Review
slides. When this report type is selected, the same options are available as for the other
reports, but the feature allowing selecting the portion of slides to print for QC Review
becomes active (no longer dimmed). (See Figure 12
6 for an example of this report). For
additional information, see Section 14.2.
•
Archive Report:
Lists all slides that were designated to No Further
Review
(See Archive
Report in Section 12.3.4 for an example of this report).
•
Process Review Report:
Lists all slides that were designated
Process Review
.
(See Figure 12
8 for an example of this report.)
•
Rerun Report:
Lists all slides that were designated
Rerun
. (See Figure 12
9 for an example
of this report.)
These reports are very similar in appearance to each other and all have the same column
headings. The arrangement and content of a report can be varied by choosing specific sorting
and layout options. The following Section 18.9.2, describes each of the sorting and layout
options in detail.
NOTE
Slide data is available to be displayed after the slides are printed
and assigned to a print set, not when they complete processing on
the instrument. Additionally, the data is presented on a 24 hour
basis, not per print set. Slides are added to the statistics the day
they are first printed and assigned to a print set, not when they
complete processing on the instrument.
Содержание FocalPoint GS
Страница 10: ...BD FocalPoint GS Imaging System Instrument User s Manual 10...
Страница 44: ...BD FocalPoint GS Imaging System Instrument User s Manual 44...
Страница 54: ...BD FocalPoint GS Imaging System Instrument User s Manual 54...
Страница 58: ...BD FocalPoint GS Imaging System Instrument User s Manual 58...
Страница 76: ...BD FocalPoint GS Imaging System Instrument User s Manual 76...
Страница 86: ...BD FocalPoint GS Imaging System Instrument User s Manual 86...
Страница 110: ...BD FocalPoint GS Imaging System Instrument User s Manual 110...
Страница 126: ...BD FocalPoint GS Imaging System Instrument User s Manual 126...
Страница 156: ...BD FocalPoint GS Imaging System Instrument User s Manual 156...
Страница 192: ...BD FocalPoint GS Imaging System Instrument User s Manual 192...
Страница 200: ...BD FocalPoint GS Imaging System Instrument User s Manual 200...
Страница 204: ...BD FocalPoint GS Imaging System Instrument User s Manual 204...
Страница 206: ...BD FocalPoint GS Imaging System Instrument User s Manual 206...
Страница 210: ...BD FocalPoint GS Imaging System Instrument User s Manual 210...
Страница 212: ...BD FocalPoint GS Imaging System Instrument User s Manual 212...
Страница 218: ...BD FocalPoint GS Imaging System Instrument User s Manual 218...
Страница 224: ...BD FocalPoint GS Imaging System Instrument User s Manual 224...