BD FACSCanto II Flow Cytometer Reference Manual
Fluid leaking from SIT or
aspirator arm
Interior valve failure or leak
1
Turn off the cytometer power.
2
Clean up the liquid, using
proper precautions.
3
Call BD Biosciences.
Waste line to fluidics cart
disconnected
1
Turn off the cytometer power.
2
Clean up the liquid, using
proper precautions.
3
Check that waste lines are
securely plugged in.
4
Turn on the cytometer power.
Instrument Troubleshooting (continued)
Observation
Possible Causes
Recommended Solutions