ELECTROSURGICAL UNIT 

ANALYZER 

 

 

 

ESU-2000A

 

 
 

USER MANUAL 

 

Содержание BC Biomedical ESU-2000A

Страница 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2000A USER MANUAL...

Страница 2: ......

Страница 3: ...SPECIFICATIONS 7 NOTES 9 BC BIOMEDICAL ESU 2000A TABLE OF CONTENTS CALIBRATION INTERVAL To ensure the accuracy of the ESU 2000A BC Group International Inc recommends that it be calibrated at least on...

Страница 4: ...ing the DUT to the ESU 2000A A serious hazard may occur if the patient is connected when testing with the ESU 2000A Do not connect any leads from the patient directly to the ESU 2000A or DUT WARNING M...

Страница 5: ...the ESU 2000A while connected to the DUT as a hazardous condition may exist CAUTION ENVIRONMENT Exposure to environmental conditions outside the specifications can adversely affect the performance of...

Страница 6: ...UAL FOR TEST PROCEDURES AND MEASUREMENT LIMITS NOTICE ABBREVIATIONS A Ampere s C Celsius Degree s DUT Device Under Test FS Full Scale Hz Hertz kg kilogram s lbs pounds M Mega 106 MHz Megahertz m milli...

Страница 7: ...MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES NOTICE CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL INC 3081 ELM POIN...

Страница 8: ...vi This Page Intentionally Left Blank...

Страница 9: ...t testing The patient return electrode resistance simulation is adjustable in one ohm increments from 0 to 999 ohms The following are highlights of some of the main features LARGE 4 INCH ANALOG DISPLA...

Страница 10: ...2 STANDARD ACCESSORIES BC20 00121 ACCESSORY KIT INCLUDES TEST LEADS ALLIGATOR CLIPS AND REPLACEMENT FUSE OPTIONAL ACCESSORIES BC20 30108 SOFT CARRYING CASE...

Страница 11: ...that are present OVERVIEW Patient Plate Return Electrode Monitor Test Terminals RF Leakage Input Terminal Oscilloscope Output Fast Blow Fuse 4 INCH ANALOG DISPLAY Potentiometer Ohms ESU Ground RF Leak...

Страница 12: ...ing does not exceed the full scale Power range selected on the ESU 2000A 7 Activate the DUT output The Analyzer will indicate both mA RMS and Power W on the appropriate color coded wattage meter range...

Страница 13: ...00A 10 Connect the single electrode REM cable to the DUT s Patient Plate or Return Electrode and the Leakage Input on the ESU 2000A 11 Set the DUT for proper output mode and power level 12 Activate th...

Страница 14: ...ENDED THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WAR...

Страница 15: ...PE OUTPUT OUTPUT Isolated Uncalibrated CONNECTION BNC 50 VARIABLE LOAD VALUES 50 100 200 300 400 500 TYPE Non Inductive ACCURACY 2 1 POWER RATING MAX 50 50 W 100 100 W 200 200 W 300 300 W 400 400 W 50...

Страница 16: ...NVIRONMENTAL DISPLAY 4 1 2 Analog Meter Color coded display W and mA CONSTRUCTION ENCLOSURE ABS Plastic FACE PLATE Back printed Lexan SIZE 5 36 x 8 68 x 9 10 Inches 136 1 x 220 5 x 231 1 mm WEIGHT 4 L...

Страница 17: ...9 NOTES...

Страница 18: ...10 NOTES...

Страница 19: ......

Страница 20: ...NTERNATIONAL INC 3081 ELM POINT INDUSTRIAL DRIVE ST CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com sales bcgroupintl com ESU 2000A User Manual 5 20 Rev 09 Copyright 2020 Made i...