9.9 Specifications
MultiGasAnalyzer
TM
MGA-3050
Physical Data
Dimensions (W x D x H)
1.49 x 1.14 x 1.22 inches
(38 x 29 x 31 mm)
Weight
< 25 g (cable excluded)
Cable length
8.2 feet ± 1 inch (2.50 m ±0.02)
Environmental Data
Operating temperature
10...40 °C (50...104 °F)
Storage temperature
–20...50 °C (-4...122 °F)
Operating humidity
10
–95% RH, non-condensing
Storage humidity
5
–100% RH, condensing
Operating atmospheric
pressure
700-1200 hPa (700 hPa corresponding
to an altitude of 3048 m)
Accuracy
specifications
Gas
Range
Accuracy
CO
2
0-10%
±(0.2% ABS + 2% REL)
10-20%
±(0.3% ABS + 4% REL)
N
2
O
0-100%
±(2% ABS + 2% REL)
HAL, ISO, ENF
0-8%
±(0.15% ABS + 5% REL)
8-12%
±(0.2% ABS + 10% REL)
SEV
0-10%
±(0.15% ABS + 5% REL)
10-15%
±(0.2% ABS + 10% REL)
DES
0-22%
±(0.15% ABS + 5% REL)
22-25%
±(0.2% ABS + 10% REL)
Rise time (@ 10
l/min)
CO
2
< 90 ms
O
2
,
N
2
O, HAL, ISO, ENF, SEV, DES < 300 ms
Monitoring
Numerical Data and Real-time Waveform with
FlowLab
TM
Software
Interference at indicated gas level. For example, 50 vol% Helium typically decreases the CO2
readings by 6%. This means that if measuring on a mixture containing 5.0 vol% CO2 and 50
vol% Helium, the measured CO2 concentration will typically be (1-0.06) * 5.0 vol% = 4.7 vol%
CO2.